Conceptualization: Kamiue M, Tsubahara A, Ito T. Methodology: Kamiue M, Tsubahara A, Ito T. Formal analysis: Kamiue M, Tsubahara A, Ito T, Koike Y. Funding acquisition: Kamiue M, Tsubahara A, Ito T. Project administration: Kamiue M, Tsubahara A, Ito T. Visualization: Kamiue M, Tsubahara A, Ito T. Writing – original draft: Kamiue M, Tsubahara A, Ito T. Writing – review and editing: Kamiue M, Tsubahara A, Ito T. Approval of final manuscript: all authors.
Coil area (cm2) | KET (Nm) | Output intensity (T) | |
---|---|---|---|
MVC | - | 161.0±51.4 | - |
A100% | 10.1 | 29.6±18.1 | 1.08 |
B70% | 191 | 54.9±23.4 | 1.07 |
B100% | 191 | 66.1±24.5 | 1.47 |
Values are presented as number only or mean±standard deviation.
The output intensity of the magnetic stimulation immediately below the probe surface was equivalent between B70% and A100%.
MVC, maximum voluntary contraction; rPMS, repetitive peripheral magnetic stimulation; KET, knee extensor torque; A100%, maximum output intensity of the stimulator manufactured by IFG Co. Ltd. (Stim A); B70%, 70% of maximum output intensity of the stimulator manufactured by REMED Co. Ltd. (Stim B); B100%, maximum output intensity of the Stim B.
Reference | Equipment | Coil properties | Stimulation frequency (Hz) | Output intensity (T) | Stimulation site | Duration of stimulation (s) | Pain assessment | Measured value of KET | Participants |
---|---|---|---|---|---|---|---|---|---|
Han et al., 2006 [27] | Biocon-1000 | Circumferential | 25 | 1.8 | Rectus femoris | 2 | VAS | 9.5±4.8 Nm | Healthy adults |
Kremenic et al., 2004 [29] | Magstim Rapid with four booster units | 110-mm double-circular | 20, 30, 40 | 2 | Main trunk of femoral nerve | 3 | None | Approximately 70% of MVC “estimated” | Healthy adults |
Kremenic et al., 2009 [30] | Magstim Rapid with eight booster units | 110-mm double-circular | 25 | 2 | Main trunk of femoral nerve | 3 | None | 89.6%±9.6% of MVC | Healthy adults |
Fukunaga et al., 2019 [31] | Magstim Rapid with eight booster units | 110-mm double-circular | 35 | 2 | Main trunk of femoral nerve | 3 | None | 168.0±79.8 Nm | Post-reconstruction of the anterior cruciate ligament |