Ann Rehabil Med Search

CLOSE


Ann Rehabil Med > Volume 44(5); 2020 > Article
Kim, Shin, Kim, Shin, Park, Kim, and Yoon: What is the Adequate Cuff Volume for Tracheostomy Tube? A Pilot Cadaver Study

Abstract

Objective

To determine the patterns of tracheostomy cuff pressure changes with various air inflation amounts in different types of tracheostomy tubes to obtain basic data for appropriately managing long-term tracheostomy.

Methods

We performed tracheostomy on a 46-year-old male cadaver. Three types of tracheostomy tubes (single-cuffed, double-cuffed, and adjustable flange), divided into 8 different subtypes based on internal tube diameters and cuff diameters, were inserted into the cadaver. Air was inflated into the cuff, and starting with 1 mL air, the cuff pressure was subsequently measured using a manometer.

Results

For the 7.5 mm/14 mm tracheostomy tube, cuff inflation with 3 mL of air yielded a cuff pressure within the recommended range of 20–30 cmH2O. The 7.5 mm/24 mm tracheostomy tube showed adequate cuff pressure at 5 mL of air inflation. Similar values were observed for the 8.0 mm/16 mm and 8.0 mm/27 mm tubes. Double-cuffed tracheostomy cuff pressures (7.5 mm/20 mm and 8.0 mm/20 mm tubes) at 3 mL air inflation had cuff pressures of 18–20 cmH2O at both the proximal and distal sites. For the adjustable flange tracheostomy tube, cuff pressure at 6 mL of cuff air inflation was within the recommended range. Maximal cuff pressure was achieved at inflation with almost 14 mL of air, unlike other tube types.

Conclusion

Various types of tracheostomy tubes showed different cuff pressures after inflation. These values might aid in developing guidelines For patients who undergo tracheostomy and are discharged home without cuff pressure manometers, this standard might be helpful to develop guidelines.

INTRODUCTION

The critical functions of endotracheal tube cuffs are blocking aspiration of pharyngeal contents into the trachea by sealing the airway and preventing air leakage through the cuff during positive pressure ventilation [1]. High-compliance endotracheal tubes are mainly used for sealing the airway to prevent pulmonary aspiration [2]. Although cuffs need to be loaded with pressures high enough to prevent the loss of tidal volume and aspiration into the trachea, problems caused by excessive cuff pressures also need to be considered [3-7]. The main issues after consistently high cuff pressures are decreased tracheal capillary perfusion, direct tracheal wall injury leading to loss of mucosal cilia, ulceration, bleeding, and even tracheoesophageal fistulas in severe cases [5,8]. Nordin [5] reported that when the cuff pressure was higher than 40 cmH2O, the cuff inflation volume transmitted direct injury to the tracheal wall. In addition, Bernhard et al. [9] reported that while a thin-walled cuff with approximately 25 cmH2O of cuff pressure is adequate in preventing tracheal aspiration, the pressure exerted by the cuff on the tracheal wall may exceed the capillary perfusion pressure when used for a longer time, resulting in ischemic damage. Various strategies have been employed to decrease tracheal injury such as regular cuff deflation, alternating inflation with double cuffs, inflating only during aspiration, and careful inflation. However, these techniques were not effective [10]. Therefore, attempts have been made to obtain a low-pressure, high-volume endotracheal tube cuff for several years. The ideal cuff inflation pressure minimizing the risk of both tracheal wall injury and aspiration has been known to be approximately 20–30 cmH2O [5,11].
The Korean National Health Insurance provides coverage for portable ventilators for patients with motor neuron diseases and respiratory failure disorders who are incapable of spontaneous breathing. A large number of these patients are released from the hospital with a tracheostomy [12]. When long-term vent care is necessary, simple cuff pressure palpation is not sensitive enough to detect problems [13]. Not many patients possess cuff pressure manometers at home. Therefore, the significance of routine cuff pressure measurement has been emphasized [14,15].
The present pilot study aimed to obtain basic data for the appropriate management of long-term tracheostomy tube application. The pressure changes in several kinds of tracheostomy tubes commonly used in clinical practice were measured and compared according to the level of cuff air inflation. Furthermore, the characteristics of each tracheostomy tube type and its prioritization for use during specific situations were also considered.

MATERIALS AND METHODS

This study conformed to the provisions of the Declaration of Helsinki. Approval from the Institutional Review Board of our institution was waived since this study did not use human subjects for research.
Tracheostomy was performed in a 46-year-old male cadaver. Three of the most commonly used tracheostomy tubes in clinical practice were then inserted and inflated with air. Cuff pressures from each were subsequently measured.
Tube types included a Portex (Smiths Medical Inc., Plymouth, MN, USA) cuffed tracheostomy tube, a Portex tracheostomy tube with a double cuff, and a Portex adjustable neck flange hyperflexion tracheostomy tube with a longer maximal cannula.
Tracheostomy tubes were divided into a total of 8 subtypes, depending on the internal tube and resting (without air inflation) cuff diameters. There were 4 subtypes for single-cuffed tracheostomy tubes, 2 subtypes for double-cuffed tracheostomy tubes, and 2 subtypes for adjustable tracheostomy tubes (Table 1).
Thereafter, air was injected slowly into the cuff of each tracheostomy tube in increments of 1 mL. To measure the pressure, a sufficient time interval was given after each milliliter (mL) of air inflation. Cuff pressure was monitored using a manometer (VBM Medizintechnik GmbH, Sulz am Neckar, Germany) until the cuff pressure exceeded 120 cmH2O. For double-cuffed tracheostomy tubes, both the proximal and distal pressures were measured. Two rehabilitation medicine clinicians measured the pressure three times each. The mean pressure value was then recorded. To ensure equivalent tracheal compliance conditions, pressures were measured by each clinician at 30-minute intervals. In addition, the pressures of different types of tubes were measured on different days.
Since it would be impossible to measure and compare the pressures of 8 types of tracheostomy tubes in a live patient, we reviewed the 7.5 mm/24 mm cuffed tracheostomy tube cuff pressures and tracheal diameters of 2 male patients from the same age group (Patient A: male, 54 years old; Patient B: male, 63 years old). These representative values were compared to those obtained with the cadaver.
For statistical analyses, the SPSS version 18 software was used to calculate the intra-class correlation coefficients (ICCs).

RESULTS

On the cadaver, tracheostomy was performed between the second and third tracheal rings [16]. After inserting the tracheostomy tube, the cuff was placed between the fifth and sixth tracheal rings. The inner tracheal diameter at the site of the cuff was 20.5 cm. On chest X-ray, the inner tracheal diameters around the cuff inflation site for the 2 patients were 20.53 mm and 19.58 mm, respectively.
The pressures showed excellent agreement between raters (ICC=0.94; 95% confidence interval, 0.87–0.98). A pressure of 20–30 cmH2O, corresponding to the green area on the manometer, was measured at 3 mL air inflation in the 7.5 mm/14 mm and 8.0 mm/16 mm cuffed tracheostomy tubes. The 7.5 mm/24 mm and 8.0 mm/27 mm tube types with bigger diameters maintained optimal pressures at 5 mL. At the same volume of inflation, all 4 types had higher pressures in tubes with smaller diameters.
The pressures in the double-cuffed tracheostomy tubes were within the recommended range in both 7.5 mm/8.0 mm tube types at 3–4 mL air inflation. No significant difference between the proximal and distal cuff pressures of the cadaver or tracheal lesions that could influence the results were observed.
The adjustable tube maintained optimal pressures at 6 mL and at 6–7 mL for the 7.0 mm/24 mm and 8.0 mm/30 mm tubes, respectively. It was possible to measure pressures up to 15 mL inflation on the manometer (Table 2).
The pressures of the 7.5 mm/24 mm cuffed tracheostomy tubes for both patients were within the recommended range at 5 mL inflation (23 cmH2O and 28 cmH2O, respectively). These were identical to the cadaveric values (Fig. 1).

DISCUSSION

In case of our cadaver, ideal air inflation for Portex cuffed tracheostomy tube and Portex tracheostomy tube with double cuff was 3–5 mL depending on the size of internal tube diameters and resting cuff diameters. The recommended representative cuff pressure for real patient showed the same result as that of the cadaver study, demonstrating the reliability of our study. For adjustable neck flange tubes, 6–7 mL was the optimal inflation volume.
To compare the characteristics depending on the type of tracheostomy tube and priority of each type of tube at specific situations, we used 3 types of the Portex cuff tracheotomy tube, which is widely used in Korea. The Portex cuff tracheotomy tube is mostly used to prevent aspiration into the trachea regardless of the patient group. It provides mild support to patients who have had a tube insertion for the first time because it is made of heat sensitive polyvinyl chloride, which becomes softer at the body temperature of patients, making it more convenient because of lower stimulation. In addition, a double-cuffed tracheotomy tube is useful for alternatively shifting the pressure between different regions of the tracheal walls. Adjustable neck flange tubes, which have a characteristic longer maximal cannula including a 15 mm connector, can be used in patients who have unusual anatomy or pathology, such as glottic edema and obesity [17].
Based on the assumption that the cuff pressure would be different depending on each individual, the present study was performed with a cadaver of a male in his 40s with no lesions in the trachea region, and with a standard tracheal lumen diameter, which on average is approximately 19.5±2.3 mm in adult males in their 40s [18]. We compared the data with those of 2 male patients of the same age and in the same tracheal lumen diameter group. In addition, the time interval was sufficient to minimize the effect of tracheal tissue transformation after formaldehyde fixation and make equivalent conditions of the trachea compliance. We selected patients who underwent insertion of the most commonly used single-cuff tracheostomy tube. In consideration of ethical concerns, we attempted to minimize tube replacement and air inflation to keep them within the recommended range for each patient. This was to determine whether the increased pressure pattern and optimal inflation volume were consistent with those of the cadaver and control groups. We confirmed that the optimal inflation volume in both the cadaver and control groups was 5 mL (Fig. 1).
Optimal capacities of air inflation and pressure were identified depending on the subtype of each tracheostomy tube. Tubes with smaller cuff diameters tended to have higher pressures at the same capacity of air inflation than tubes of the same type with bigger diameters. However, it could not be simply concluded that a bigger cuff is more advantageous, because we were unable to determine if a tube with a bigger cuff would have a better sealing effect than a tube with smaller one at the corresponding air inflation. To this end, in cases of real patients, air leakage would have to be checked through mechanical ventilation.
In addition, the present study found no significant difference between proximal and distal cuff pressures of double-cuffed tubes. However, it is thought that alternative ballooning of double-cuffed tubes would be useful for minimizing the degree of damage in cases of tracheal stenosis, tracheomalacia, and focal tracheal lesion due to a long-term tracheostomy [19].
An additional remarkable result was regarding the cuff pressure of the adjustable tubes. It was easily predictable that the tube with the biggest resting cuff diameter (30 mm) in this tube type would have a higher maximal air inflation capacity within the recommended pressure range than the smaller resting cuff diameter tubes. However, the pressure was lower in adjustable tubes than in single-cuffed tracheostomy tube at the same capacity of inflation compared to single-cuffed tubes with the same resting diameter when injecting the same volume, which is thought to be related with normal tracheal anatomy (Fig. 2). In other words, since it is natural that the diameter and cross-sectional area increase in the portion of the trachea close to the carina, an adjustable tube with a longer tracheal portion than that of other tubes is inserted deeper, so that the pressure between the tracheal wall for a broader area and the cuff is measured with an adjustable tube compared to other types, resulting in the above results [20].
Several limitations of the present study should be considered. First, the results from a cadaver cannot be directly applied to patients. If there are insufficient intervals while performing tracheostomy ballooning several times in a cadaver fixed with formaldehyde, inter-rater and intra-rater reliability of pressure measurements could be affected by already stiffened smooth muscle. Therefore, the present study attempted to enhance the reliability of data by measuring at different days and at 30-minute intervals. In addition, this was a preliminary study conducted on a male cadaver in his 40s using 2 patients as controls. To present standardized data, further studies accounting for differences in sex, age, and tracheal diameter are necessary. In our study, considering the possible difference in compliance of the tracheal tissue between the cadaver and the patients, it was meaningful to confirm that the air inflation volume within the optimal pressure was the same (Fig. 1) for both groups. Second, as high volume and low pressure were used in this study, there is a possibility of the cuff inflation volume and pressure increasing owing to the body temperature [21]. Further studies are needed to evaluate the effects of body temperature on changes in pressure pattern. Third, we did not compare the different effects of texture of the cuff membrane. Dullenkopf et al. [22] performed an in vitro study that showed that the ultrathin polyurethane cuff was desirable for high volume and low pressure compared to other types. Since the results were obtained through a cadaver, additional experimental groups would be needed to generalize these values for direct application to patients. However, this study was meaningful to emphasize the importance of routine screenings and necessity for monitoring adequate cuff pressure for long-term tracheostomy state patients.
In conclusion, tracheostomy tubes showed different cuff pressures after inflation characteristics depending on the type of tracheostomy tube. For tracheostomy state patients who are discharged home without cuff pressure manometers, this standard might be helpful to develop guidelines. In addition, double-cuffed tracheostomy using adjustable flange tracheostomy tubes can be recommended for some patients requiring more pressure.

CONFLICTS OF INTEREST

No potential conflict of interest relevant to this article was reported.

ACKNOWLEDGMENTS

This work was supported by clinical research grant from Pusan National University Hospital in 2020.

AUTHOR CONTRIBUTION

Conceptualization: Kim DM, Yoon JA. Methodology: Kim DM, Yoon JA. Formal analysis: Yoon JA, Shin YB. Funding acquisition: Yoon JA. Project administration: Yoon JA, Shin YB. Visualization: Yoon JA, Kim SD. Writing – original draft: Kim DM, Yoon JA. Writing – review and editing: Kim DM, Yoon JA, Shin MJ, Park HE, Kim YM. Approval of final manuscript: all authors.

Fig. 1.
Equivalent air inflation (5 mL) for ideal cuff pressure (20–30 cmH2O) in the cadaver and 2 patients with 7.5 mm/24 mm tube insertion (Portex 100/518/075).
arm-19210f1.jpg
Fig. 2.
Lower cuff pressure with equal air inflation in an adjustable tracheostomy tube (100/523/070) compared to a single-cuffed tracheostomy tube (Portex 100/518/075).
arm-19210f2.jpg
Table 1.
Types of tracheostomy tubes
Tube type Tube subtype (catalog#) Internal tube diameter (mm) Resting cuff diameter (mm)
Portex cuffed tracheostomy tube 100/510/075 7.5 14
100/510/080 8.0 16
100/518/075 7.5 24
100/518/080 8.0 27
Portex tracheostomy tube with double cuff 100/512/075 7.5 20, double cuff
100/512/080 8.0 20, double cuff
Portex cuffed adjustable flange tracheostomy tube 100/523/070 7.0 24, adjustable
100/523/080 8.0 30, adjustable
Table 2.
Tracheostomy cuff pressure of cadaver for each type of tube after inflation
Volume (mL) Cuff pressure (cmH2O)
100/510/075 100/512/075
100/518/075 100/510/080 100/512/080
100/518/080 100/523/070 100/523/080
Proximal Distal Proximal Distal
1 Null 2.1± 0.1 2.0±0.0 Null 3.5±0.0 Null Null Null Null Null
2 2.3±0.0 9.5±0.0 8.0±0.0 1.3±0.1 14.4±0.2 4.2±0.2 3.0±0.0 Null Null Null
3 26.4±0.1 18.2±0.2 17.2±0.2 2.2±0.2 20.4±0.2 19.4±0.2 17.0±0.0 2.1±0.3 5.0±0.0 6.0±0.0
4 62.0±0.0 40.5±0.0 41.0±0.0 10.3±0.1 38.2±0.0 50.3±0.1 51.2±0.1 10.2±0.2 8.0±0.0 12.0±0.0
5 60.3±0.1 74.2±0.3 76.0±0.0 24.3±0.3 97.3±0.1 66.0±0.0 65.5±0.1 36.5±0.0 12.2±0.1 20.5±0.1
6 66.5±0.0 116.4±0.2 114.5±0.0 38.5±0.0 114.3±0.2 98.0±0.0 97.5±0.0 38.0±0.0 18.2±0.1 24.0±0.0
7 >120 >120 >120 44.4±0.1 >120 >120 >120 44.0±0.0 40.5±0.0 30.0±0.0
8 - - - 72.7±0.2 - - - 58.5±0.0 62.5±0.0 38.5±0.0
9 - - - 96.4±0.1 - - - 86.0±0.0 88.0±0.0 44.2±0.3
10 - - - >120 - - - >120 100.0±0.0 60.5±0.0
11 - - - - - - - - >120 78.7±0.3
12 - - - - - - - - - 88.5±0.0
13 - - - - - - - - - 110.5±0.1
14 - - - - - - - - - 118.0±0.0
15 - - - - - - - - - >120

REFERENCES

1. Sengupta P, Sessler DI, Maglinger P, Wells S, Vogt A, Durrani J, et al. Endotracheal tube cuff pressure in three hospitals, and the volume required to produce an appropriate cuff pressure. BMC Anesthesiol 2004;4:8.
crossref pmid pmc pdf
2. Braz JR, Navarro LH, Takata IH, Nascimento Junior P. Endotracheal tube cuff pressure: need for precise measurement. Sao Paulo Med J 1999;117:243-7.
crossref pmid pdf
3. Stauffer JL, Olson DE, Petty TL. Complications and consequences of endotracheal intubation and tracheotomy: a prospective study of 150 critically ill adult patients. Am J Med 1981;70:65-76.
crossref pmid
4. Leigh JM, Maynard JP. Pressure on the tracheal mucosa from cuffed tubes. Br Med J 1979;1:1173-4.
crossref pmid pmc
5. Nordin U. The trachea and cuff-induced tracheal injury: an experimental study on causative factors and prevention. Acta Otolaryngol Suppl 1977;345:1-71.
pmid
6. Klainer AS, Turndorf H, Wu WH, Maewal H, Allender P. Surface alterations due to endotracheal intubation. Am J Med 1975;58:674-83.
crossref pmid
7. Shelly WM, Dawson RB, May IA. Cuffed tubes as a cause of tracheal stenosis. J Thorac Cardiovasc Surg 1969;57:623-7.
crossref pmid
8. Bernhard WN, Yost L, Joynes D, Cothalis S, Turndorf H. Intracuff pressures in endotracheal and tracheostomy tubes. Related cuff physical characteristics. Chest 1985;87:720-5.
crossref pmid
9. Bernhard WN, Cottrell JE, Sivakumaran C, Patel K, Yost L, Turndorf H. Adjustment of intracuff pressure to prevent aspiration. Anesthesiology 1979;50:363-6.
crossref pmid pdf
10. Bryant LR, Trinkle JK, Dubilier L. Reappraisal of tracheal injury from cuffed tracheostomy tubes: experiments in dogs. JAMA 1971;215:625-8.
crossref pmid
11. Park YJ, Lee J, Kim SH, Ko SH, Shin MJ, Chang JH, et al. Care status of the ALS patients with long-term use of tracheostomy tube. Ann Rehabil Med 2015;39:964-70.
crossref pmid pmc pdf
12. Suh MR, Choi WA, Choi YC, Lee JW, Hong JH, Park J, et al. Long-Term outcome of amyotrophic lateral sclerosis in Korean subjects. Ann Rehabil Med 2017;41:1055-64.
crossref pmid pmc pdf
13. Fernandez R, Blanch L, Mancebo J, Bonsoms N, Artigas A. Endotracheal tube cuff pressure assessment: pitfalls of finger estimation and need for objective measurement. Crit Care Med 1990;18:1423-6.
crossref pmid
14. Dunn CR, Dunn DL, Moser KM. Determinants of tracheal injury by cuffed tracheostomy tubes. Chest 1974;65:128-35.
crossref pmid
15. Lewis FR Jr, Schiobohm RM, Thomas AN. Prevention of complications from prolonged tracheal intubation. Am J Surg 1978;135:452-7.
crossref pmid
16. Durbin CG Jr. Techniques for performing tracheostomy. Respir Care 2005;50:488-96.
pmid
17. Hess DR. Tracheostomy tubes and related appliances. Respir Care 2005;50:497-510.
pmid
18. Breatnach E, Abbott GC, Fraser RG. Dimensions of the normal human trachea. AJR Am J Roentgenol 1984;142:903-6.
crossref pmid
19. St John RE, Malen JF. Contemporary issues in adult tracheostomy management. Crit Care Nurs Clin North Am 2004;16:413-30.
crossref pmid
20. Otoch JP, Minamoto H, Perini M, Carneiro FO, de Almeida Artifon EL. Is there a correlation between right bronchus length and diameter with age? J Thorac Dis 2013;5:306-9.
pmid pmc
21. Al-Shaikh B, Stacey SG. Essentials of equipment in anaesthesia, critical care, and peri-operative medicine. Edinburgh, UK: Elsevier; 2018.

22. Dullenkopf A, Gerber A, Weiss M. Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube. Intensive Care Med 2003;29:1849-53.
crossref pmid pdf
TOOLS
Share :
Facebook Twitter Linked In Line it
METRICS Graph View
  • 1 Web of Science
  • 1 Crossref
  • 2 Scopus
  • 3,699 View
  • 95 Download
Related articles in ARM


ABOUT
ARTICLE TYPES

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Terms of Use   |   Privacy Polity
Editorial Office
Department of Rehabilitation Medicine, Seoul National University Hospital
101 Daehak-ro, Jongno-gu, Seoul, Korea
Tel: +82-10-8678-2671    Fax: +82-2-6072-5244    E-mail: edit@e-arm.org; edit.karm@gmail.com
Business Registration: 110-82-07460                

Copyright © 2024 by Korean Academy of Rehabilitation Medicine.

Developed in M2PI

Close layer